Patient Monitoring
Patient Monitor · NIBP
Private healthcare facility
Problem
Inaccurate blood pressure readings on the patient monitor.
Solution
Detected a fine leak in the NIBP cuff and repaired the leak source.
Result
Measurement accuracy restored within manufacturer specifications.